Título: Comparação entre os tonômetros portáteis Tonovet, Tonovet Plus, Tono-Pen Avia Vet e Kowa HA-2 na mensuração da pressão intraocular em cavalos
Título(s) alternativo(s): Comparison between Tonovet, Tonovet Plus, Tono-Pen Avia Vet and Kowa HA-2 portable tonometers in the measurement of intraocular pressure in horses
Autor: Angeluci, Gismelli Cristiane 
Primeiro orientador: Andrade, Silvia Maria Caldeira Franco
Primeiro membro da banca: Zulim, Luis Felipe da Costa
Segundo membro da banca: Souza, Ana Letícia Groszewicz de
Resumo: O objetivo do presente estudo foi comparar e avaliar a acurácia dos tonômetros portáteis com diferentes metodologias, Tonovet, Tonovet Plus, Tono-Pen Avia Vet e Kowa HA-2 na mensuração da pressão intraocular (PIO) em cavalos. A mensuração da PIO tem importância para o diagnóstico de oftalmopatias que podem levar à cegueira irreversível do animal, como aquelas que podem levar ao aumento da PIO, como o glaucoma ou à sua diminuição, como as uveítes. O diagnóstico é realizado principalmente com o uso de tonômetros que aferem a PIO, sendo a tonometria de aplanação e rebote as mais utilizadas atualmente na Medicina Veterinária representados pelos tonômetros portáteis Tono-Pen Avia Vet (aplanação), Tonovet (rebote), mais recentemente o Tonovet Plus (rebote) e Kowa Ha-2 (metodologia de Goldmann). Foram avaliados 12 olhos de 6 cavalos sedados, para o estudo in vivo comparando a PIO real (manometria) com a tonometria, e 60 olhos sadios de 30 cavalos hígidos não sedados, em um estudo a campo com a PIO mensurada com diferentes tonômetros. No estudo in vivo os valores da PIO em mmHg mensurados na manometria ocular foram: 24,9±4,0 (20,0-30,0) e na tonometria: Tonovet 25,7±5,8 (19,5-33,0), Tonovet Plus 24,8±7,1 (13,2-33,2), Tono-Pen Avia Vet 19,2±4,7 (13,1-26,5), Kowa HA-2 24,1±1,2 (22,8-25,8); no estudo a campo os valores da PIO foram: Tonovet 30,7±5,6 (21,7-38,0), Tonovet Plus 29,6±6,7 (16,2-38,6), Tono-Pen Avia Vet 27,3±5,8 (14,6-37,1), Kowa HA-2 23,4±2,2 (20,2-28,7). Foi verificada uma forte correlação entre os valores da PIO com a manometria em todos os tonômetros. Valores mais altos da PIO foram aferidos com o Tonovet Plus e Tonovet e valores mais baixos com o Tono-Pen Avia Vet e o Kowa HA-2. Todos os tonômetros utilizados tiveram precisão na mensuração da PIO dos cavalos, especificamente o tonômetro mais recente, o Tonovet Plus, que mostrou uma excelente correlação com a manometria.
Abstract: The aim of the present study was to compare and evaluate portable tonometers accuracy with different methodologies, Tonovet, Tonovet Plus, Tono-Pen Avia Vet and Kowa HA-2 in measuring IOP in horses. Measurement of intraocular pressure (IOP) is important for ophthalmopathies diagnosis that can lead to irreversible animal blindness, such as those lead to IOP increase, in glaucoma or its decrease, in uveitis. Diagnosis is performed mainly with tonometers that measure the IOP, applanation and rebound tonometry are the most used nowadays in Veterinary Medicine, represented by the portable tonometers Tono-Pen Avia Vet (applanation), Tonovet (rebound), more recently Tonovet Plus (rebound) and Kowa Ha-2 (Goldmann methodology). Twelve eyes from 6 sedated horses were used for in vivo study comparing actual IOP (manometry) with tonometry, and 60 healthy eyes from 30 healthy unsedated horses in a field study with IOP measured with different tonometers. In vivo study, the IOP values in mmHg measured in ocular manometry were: 24.9±4.0 (20.0-30.0) and in tonometry: Tonovet 25.7±5.8 (19.5-33 .0), Tonovet Plus 24.8±7.1 (13.2-33.2), Tono-Pen Avia Vet 19.2±4.7 (13.1-26.5), Kowa HA-2 24 0.1±1.2 (22.8-25.8); in field study the IOP values were: Tonovet 30.7±5.6 (21.7-38.0), Tonovet Plus 29.6±6.7 (16.2-38.6), Tono-Pen Avia Vet 27.3±5.8 (14.6-37.1), Kowa HA-2 23.4±2.2 (20.2-28.7). There was a strong correlation between IOP values and manometry in all tonometers. Higher IOP values were measured with Tonovet Plus and Tonovet and lower values with Tono-Pen Avia Vet and Kowa HA-2. All tonometers used were accurate in measuring the IOP of horses, specifically the most recent tonometer, the Tonovet Plus, which showed an excellent correlation with manometry.
Palavras-chave: tonometria, manometria ocular, tonometria de rebote, tonometria de aplanação, Tonovet Plus, cavalo.
tonometry, ocular manometry, rebound tonometry, applanation tonometry, Tonovet Plus, horse.
Área(s) do CNPq: CIENCIAS AGRARIAS::MEDICINA VETERINARIA
Idioma: por
País: Brasil
Instituição: Universidade do Oeste Paulista
Sigla da instituição: UNOESTE
Departamento: Mestrado em Ciência Animal
Programa: Mestrado em Ciência Animal
Citação: Angeluci, Gismelli Cristiane. Comparação entre os tonômetros portáteis Tonovet, Tonovet Plus, Tono-Pen Avia Vet e Kowa HA-2 na mensuração da pressão intraocular em cavalos. 2021. 50f. Dissertação (Mestrado em Ciência Animal) - Universidade do Oeste Paulista, Presidente Prudente, 2021.
Tipo de acesso: Acesso Aberto
URI: http://bdtd.unoeste.br:8080/jspui/handle/jspui/1408
Data de defesa: 7-Out-2021
Aparece nas coleções:Mestrado em Ciência Animal

Arquivos associados a este item:
Arquivo Descrição TamanhoFormato 
Gismelli Cristiane Angeluci.pdfGismelli Cristiane Angeluci917,55 kBAdobe PDFThumbnail

Baixar/Abrir Pré-Visualizar


Os itens no repositório estão protegidos por copyright, com todos os direitos reservados, salvo quando é indicado o contrário.